-
iestinstrument
Correlating Electrode Tortuosity and McMullin Number with Battery Rate Performance
Abstract
Electrode tortuosity (\(\tau\)) is a dimensionless microstructural parameter quantifying how much the actual ionic transport path through a porous battery electrode deviates from a straight line; the MacMullin number (McMullin number, \(N_m = \tau / \varepsilon\)) combines tortuosity and porosity into a single EIS-measurable metric that directly predicts ionic resistance and rate capability: a lower MacMullin number indicates a more permeable electrode and better high-rate performance, calculated as \(N_m = R_{ion} \times A / (\sigma \times d)\) from symmetric cell impedance data. This article quantifies electrode tortuosity and MacMullin number for graphite anodes of three coating thicknesses via symmetric cell EIS, confirming a clear trend: as thickness increases from 100 \(\mu\)m to 400 \(\mu\)m, \(N_m\) rises from 4.61 to 6.15 to 6.61 and rate capacity retention degrades correspondingly — demonstrating that tortuosity testing can rank electrode rate performance without long-duration cycling.
1. Introduction
Improving cell energy density motivates the development of thicker electrodes. However, as electrode thickness increases, liquid-phase lithium-ion transport through the porous coating becomes progressively more restricted — increasing internal resistance, reducing active material utilization, and degrading both cycle performance and rate capability[1].
Electrode tortuosity quantifies the degree of curvature of the ionic transport pathway through the porous electrode structure, it is another important parameter related to the transmission characteristics besides porosity[2]. Beyond porosity (\(\varepsilon\)), tortuosity (\(\tau\)) is the critical microstructural variable governing ionic transport: two electrodes with identical porosity but different pore connectivity and geometry can have dramatically different tortuosity values and therefore very different rate performance. The MacMullin number (\(N_m = \tau / \varepsilon\)), combines both parameters into a single experimentally accessible metric that characterizes the effective ionic permeability of an electrode coating and predicts rate capability without requiring extended cell cycling.
Researchers conventionally evaluate electrode rate and cycle performance by assembling coin cells or pouch cells — but the long testing cycles required for this approach limit R\&D throughput. Electrode tortuosity testing via symmetric cell EIS provides a rapid, non-cycling alternative: a single EIS measurement can rank electrodes by their ionic transport efficiency in hours rather than days, enabling systematic screening of electrode formulation, compaction density, coating thickness, and manufacturing process variables.
2. Test Conditions and Methods
2.1 Test Equipment
- Symmetric cell assembly and EIS testing: The IEST EIC1400 Multi-Channel Ion Conductivity Test System as shown in Figure 1, integrating four-channel symmetric cell assembly fixtures, electrochemical impedance test, fitting software, and a high-purity argon atmosphere — enabling multi-channel rapid EIS measurement for electrode tortuosity determination. Frequency range: 0.1–1,000 Hz.
- Half-cell rate testing: 2032 coin cells with graphite anodes (three thickness variants) vs. lithium metal, cycled at 0.1C, 0.2C, and 0.5C to measure rate-dependent capacity and capacity retention.
Figure 1. IEST EIC1400 Multi-Channel Ion Conductivity Testing System — integrating symmetric cell assembly, argon atmosphere, and EIS measurement for electrode tortuosity and MacMullin number determination.
2.2 Electrode Preparation
Graphite slurry was coated at three doctor-blade gap settings to produce electrodes nominally 100 µm, 200 µm, and 400 µm thick (wet gap control). All electrodes were dried and handled under controlled conditions to ensure comparable porosity and binder distribution across the three thickness variants — isolating thickness as the primary variable.
2.3 Testing Protocol
- Electrode tortuosity measurement: Stack electrode–separator–electrode in each of the four channels → seal chamber, evacuate and backfill with high-purity argon → inject electrolyte quantitatively into each channel → equilibrate 10 min → measure EIS spectrum → extract Rion from Nyquist plot → compute MacMullin number using Equation (2).
- Rate performance test: Charge and discharge at 0.1C, 0.2C, and 0.5C for each electrode thickness variant; record capacity and capacity retention at each rate.
2.4 McMullin Number Calculation Method
The ionic resistance \(R_{ion}\) of the electrode coating is related to tortuosity \(\tau\), porosity \(\varepsilon\), electrolyte conductivity \(\sigma\), electrode thickness \(d\), and electrode area \(A\) by Equation (1):
\[
\tau = \frac{R_{ion} \cdot A \cdot \varepsilon \cdot \sigma}{d} \tag{1}
\]
Since direct measurement of electrode porosity \(\varepsilon\) requires complex and time-consuming characterization, the MacMullin number (also written McMullin number) \(N_m = \tau / \varepsilon\) is used instead as the standard metric for electrode tortuosity assessment. \(N_m\) is calculated directly from EIS data using Equation (2), requiring only \(R_{ion}\), electrode geometry, and electrolyte conductivity:
\[
N_m = \frac{\tau}{\varepsilon} = \frac{R_{ion} \cdot A \cdot \sigma}{d} \tag{2}
\]
The impedance of the symmetric cell was tested using the electrochemical workstation, and the EIS obtained is shown in Figure 2. Which shows a characteristic two-segment response — a high-frequency line segment and a low-frequency line segment — reflecting the ionic transport signature in the absence of electrochemical reactions. The ionic resistance \(R_{ion}\) is extracted by extending the low-frequency line segment until it intersects the real axis, then subtracting the high-frequency real-axis intercept: \(R_{ion} = \Delta Z_{Re} / 3\). Substituting into Equation (2) yields the MacMullin number, which can be analyzed as the degree of the zigzagging of the electrode.
Figure 2. Nyquist plot of symmetric cell EIS for electrode tortuosity measurement. Ionic resistance \(R_{ion}\) is extracted as \(\Delta Z_{Re} / 3\) from the intercept difference between high-frequency and extrapolated low-frequency line segments.
3. Results Analysis
3.1 MacMullin Number Results vs. Electrode Thickness
EIS fitting of the three graphite electrode variants yielded the following MacMullin numbers (Figure 3a–d):
| Electrode Thickness | MacMullin Number (Nm) | Rate Capability Rank | Interpretation |
|---|---|---|---|
| 100 µm | 4.61 | Best | Lowest electrode tortuosity; most direct ionic transport pathways; highest capacity retention at all test rates |
| 200 µm | 6.15 | Intermediate | Moderate electrode tortuosity; increased transport distance reduces high-rate utilization |
| 400 µm | 6.61 | Worst | Highest electrode tortuosity; greatest deviation from straight-path transport; most degraded rate performance |
The trend is clear: increasing electrode thickness increases the MacMullin number, indicating more tortuous ion pathways and reduced effective ionic conductivity.
Figure 3. Symmetric cell EIS Nyquist plots and MacMullin number (McMullin number) results: (a) 100 µm, Nm = 4.61; (b) 200 µm, Nm = 6.15; (c) 400 µm, Nm = 6.61; (d) MacMullin number comparison — confirming that electrode tortuosity increases monotonically with coating thickness.
The trend is physically intuitive: as electrode thickness increases, the pore network through which lithium ions must transport becomes longer and more complex. Blind pores, narrow pore throats, and tortuous inter-particle channels — each contributing to electrode tortuosity — become statistically more prevalent and more consequential over greater coating depths. The result is that ionic resistance scales faster than linearly with thickness, producing MacMullin numbers that increase from 4.61 (100 µm) to 6.61 (400 µm) — a 43% increase for a 4× thickness increase.
3.1.1 Electrode Pore Structure and Its Relationship to Tortuosity
Electrode pore size and pore connectivity are the microstructural determinants of electrode tortuosity. Porous battery electrodes contain three types of pore features that each contribute to the MacMullin number: (1) through-pores — pores that span the full electrode thickness and provide direct ionic transport pathways; these minimize electrode tortuosity; (2) dead-end (blind) pores — pores connected to the electrode surface on only one side; ions enter but must reverse direction to continue transport, increasing effective path length and tortuosity; (3) constricted pore throats — narrow junctions between pore bodies that restrict ionic flux and increase local resistance, amplifying the MacMullin number disproportionately to their volume fraction. As electrode thickness increases, the probability of encountering blind pores and constricted throats over the transport path increases — explaining why the 400 µm electrode has a significantly higher MacMullin number (6.61) than the 100 µm electrode (4.61), despite comparable porosity.
3.2 Two Methods for Electrode Tortuosity Measurement
Figure 4. Electrochemical methods for electrode tortuosity determination: (a) eRDM Polarization-Interrupt Method — MacMullin number from potential relaxation slope; (b) eSCM Symmetric Cell EIS Method — MacMullin number from ionic resistance fitting on Nyquist plot.[3]
Two standard electrochemical methods are used to determine electrode tortuosity and McMullin number:
- Polarization-Interrupt Method (eRDM): a fixed DC current is applied to polarize the cell, creating a lithium-ion concentration gradient. When the current is interrupted, the gradient relaxes and the potential decays to zero. A semi-log plot of potential vs. time is fitted to extract the MacMullin number from the relaxation curve slope.
- Symmetric Cell EIS Method (eSCM): the electrode is assembled in a symmetric non-intercalating cell, EIS is measured, and Rion is extracted from the Nyquist plot to calculate the MacMullin number via Equation (2). This is the method used in this study and implemented in the IEST EIC1400 system.
Figure 5. Critical comparison of eRDM and eSCM electrode tortuosity methods for four electrodes with identical porosity but different pore geometries: eRDM produces identical MacMullin numbers for all four (insensitive to pore structure); eSCM correctly differentiates their electrode tortuosity — demonstrating that the symmetric cell EIS method more accurately reflects actual ionic transport performance.[3]
Figure 5 demonstrates a key limitation of the eRDM method: for electrodes with identical porosity but different pore connectivity, eRDM produces identical MacMullin numbers regardless of pore geometry — because it is sensitive only to bulk concentration gradients, not to local pore structure. The eSCM method resolves these differences, producing distinct McMullin numbers that reflect actual ionic transport efficiency. The IEST EIC1400 system implements the eSCM method and therefore provides electrode tortuosity data that more faithfully represents real battery rate performance.
3.3 Correlation Between MacMullin Number and Rate Performance
Figure 6 and Table 2 demonstrate the capacity and capacity retention of negative electrode sheets with different thicknesses at different multiplicities. It can be seen that the capacity of each electrode decreases with the increase of multiplicity, but the capacity retention rate is 100 μm > 200 μm > 400 μm, which indicates that the electrode obtained by 100 μm coating has the best multiplicity performance, and the 400 μm electrode has the worst performance. Combined with the data of McMullin number, it can be seen that with the increase of the thickness of the electrode, the curvature of the electrode becomes larger, and the multiplicity performance of the cell becomes worse subsequently.
Figure 6. Rate performance of graphite anodes at three electrode thicknesses: (a) capacity-voltage profiles at 0.1C, 0.2C, 0.5C; (b) rate capacity retention curves — retention order: 100 µm (Nm = 4.61) > 200 µm (Nm = 6.15) > 400 µm (Nm = 6.61), directly correlating with electrode tortuosity.
| Rate | 100 μm | 200 μm | 400 μm | |||
|---|---|---|---|---|---|---|
| Charge (mAh/g) | Discharge (mAh/g) | Charge (mAh/g) | Discharge (mAh/g) | Charge (mAh/g) | Discharge (mAh/g) | |
| 0.1C | 259.713 | 260.862 | 225.379 | 227.152 | 175.783 | 175.084 |
| 0.2C | 222.513 | 224.121 | 191.696 | 193.351 | 139.694 | 139.111 |
| 0.5C | 136.860 | 137.320 | 98.330 | 98.448 | 57.428 | 57.312 |
The electrochemical performance test results and the pore curvature test results can completely correspond to each other, which indicates that we can predict the performance of the electrode through the test of the electrode pore curvature, correlate the electrode structure and performance prediction quickly, and accelerate the design and process development of the electrode.
4. Summary
Symmetric cell EIS measurement of electrode tortuosity on graphite anodes of three coating thicknesses confirms that electrode tortuosity increases with electrode thickness — from Nm = 4.61 at 100 µm to Nm = 6.61 at 400 µm — and that this tortuosity increase corresponds directly to degraded rate capability across all test C-rates.
Key findings: electrode tortuosity, quantified as the MacMullin number (\(N_m = \tau / \varepsilon\)), increases from 4.61 to 6.61 as graphite anode thickness increases from 100 \(\mu\)m to 400 \(\mu\)m — a 43\% increase in electrode tortuosity for a 4× increase in thickness, reflecting the greater prevalence of dead-end pores and constricted pore throats in thicker coatings. Rate capacity retention order at all test rates (0.1C, 0.2C, 0.5C) matches the MacMullin number ranking exactly, confirming that symmetric cell EIS tortuosity testing can rank electrode rate performance without cell cycling. Beyond electrode thickness, the same EIC1400 EIS method applies to screening electrode formulation (binder, carbon black), porosity (compaction density), active material morphology, electrolyte type, and separator choice — providing a rapid microstructure-to-performance correlation tool for battery electrode development.
5. References
[1] Sun Weibing et al. A low curvature thick electrode and its preparation method and application. CN115312777A. 2022.
[2] Wang Chenyang, Zhang Anbang, Chang Zenghua, et al. Progress in structure design and preparation of porous electrodes for lithium ion batteries[J]. Materials Engineering, 2022, 50 (1): 67-79.
[3] Tuan-Tu Nguyen, Arnaud Demortière, Benoit Fleutot, Bruno Delobel, Charles Delacourt & Samuel J. Cooper. The electrode tortuosity factor: why the conventional tortuosity factor is not well suited for quantifying transport in porous Li-ion battery electrodes and what to use instead [J]. Electrochem. Soc. 2018,165:A388.
6. FAQs
6.1 What is electrode tortuosity in lithium-ion batteries?
Electrode tortuosity (τ) is a dimensionless parameter that quantifies the degree to which lithium-ion transport paths through a porous battery electrode deviate from a straight line. A tortuosity of τ = 1 would mean perfectly straight pores; real electrodes have τ > 1 because pores wind, branch, narrow at junctions, and include dead-end segments that force ions to reverse direction. Higher electrode tortuosity means longer effective transport distances, higher ionic resistance, and degraded rate performance — particularly in thick electrodes where transport distance compounds the effect of tortuous pathways. Electrode tortuosity is measured electrochemically using symmetric cell EIS (the eSCM method) or the polarization-interrupt method (eRDM), with the eSCM method providing more accurate discrimination between electrodes of different pore geometries but similar porosity.
6.2 What is the MacMullin number (McMullin number) and how is it calculated?
The MacMullin number (also written McMullin number, Nm) is defined as Nm = τ / ε, the ratio of electrode tortuosity (τ) to electrode porosity (ε). It is the standard dimensionless metric for characterizing the effective ionic transport resistance of a porous battery electrode — a higher MacMullin number means more tortuous ion pathways and higher ionic resistance. Nm is calculated from symmetric cell EIS data using the equation Nm = Rion × A / (σ × d), where Rion is the ionic resistance extracted from the Nyquist plot, A is electrode area, σ is electrolyte conductivity, and d is electrode thickness. For graphite anodes measured in this study: 100 µm electrode Nm = 4.61, 200 µm Nm = 6.15, 400 µm Nm = 6.61. Lower MacMullin numbers indicate better ionic transport and predict higher rate capability.
6.3 How does electrode pore size and pore structure affect electrode tortuosity?
Electrode pore size, shape, and connectivity are the microstructural determinants of electrode tortuosity and MacMullin number. Three types of pore features determine how tortuous ionic transport becomes: (1) through-pores spanning the full electrode thickness minimize tortuosity by providing direct transport pathways; (2) dead-end (blind) pores increase effective path length because ions must reverse direction to exit — raising tortuosity significantly; (3) constricted pore throats (narrow junctions between larger pore bodies) restrict flux and amplify local resistance beyond their volume fraction. As electrode thickness increases, all three features become more consequential over the total transport distance — which is why a 4× increase in graphite anode thickness (100 µm to 400 µm) produces a 43% increase in MacMullin number (4.61 to 6.61) and measurably degraded rate capability. Electrode formulation and manufacturing parameters (compaction density, calendering, binder content, particle morphology) all affect pore geometry and therefore electrode tortuosity.
6.4 How is electrode tortuosity measured by impedance spectroscopy?
Electrode tortuosity is measured by impedance spectroscopy using a symmetric cell assembled with two identical electrode samples facing each other (electrode–separator–electrode), filled with electrolyte and left to equilibrate for 10 minutes before EIS measurement. The resulting Nyquist plot shows two characteristic line segments: a high-frequency segment (x-axis intercept = 2 × contact resistance) and a low-frequency segment that extends toward zero at −45°. The ionic resistance Rion of the electrode coating is extracted as one-third of the difference between the x-axis intercepts of the extrapolated low-frequency and high-frequency segments. Substituting Rion into Nm = Rion × A / (σ × d) gives the McMullin number. The IEST EIC1400 system automates this workflow with argon-atmosphere cell assembly, four-channel simultaneous measurement, and integrated fitting software.
6.5 Why does electrode tortuosity increase with coating thickness?
Electrode tortuosity increases with coating thickness because the total transport path length through which lithium ions must navigate a tortuous pore network grows faster than linearly with thickness. In a thin electrode (100 µm), most ions can find relatively direct pathways through the pore structure; the cumulative effect of dead-end pores and pore constrictions is limited by the short total distance. In a thick electrode (400 µm), ions must traverse a much longer pore network, encountering dead-end pores, branching junctions, and constricted throats more frequently — each deviation from a straight path adds to the effective transport distance and increases Rion. In this study, a 4× increase in graphite anode thickness produced a 43% increase in MacMullin number (Nm = 4.61 to Nm = 6.61), corresponding to measurably degraded rate capacity retention at all tested C-rates.
6.6 What variables besides thickness can be screened using electrode tortuosity testing?
Symmetric cell EIS electrode tortuosity testing and MacMullin number measurement can screen any electrode design or process variable that affects pore structure and ionic transport, including: electrode formulation (binder type and content, carbon black loading, active material particle size distribution and morphology); compaction density and calendering pressure; coating uniformity and solvent system; electrolyte type and viscosity; separator type (which contributes its own tortuosity to the cell EIS signal); and aging state (comparing fresh vs. cycled electrodes to detect pore-structure changes from swelling, cracking, or SEI growth). The IEST EIC1400 four-channel system enables parallel measurement of multiple variants in a single test session — accelerating electrode formulation screening compared to the days required for rate performance cycling in half-cells.
Contact Us
If you are interested in our products and want to know more details, please leave a message here, we will reply you as soon as we can.







