- Home
- Battery Slurry Resistivity Case
- A Method for Characterizing Slurry Resistivity and Sedimentation
Quantitative Analysis of the Swelling of Different Proportions of Silicon Carbon Negative Electrode
Literature appreciation: Electrochemical Dilatometry of Si-Bearing Electrodes: Dimensional Changes and Experiment Design
1. Author Information and A Summary of the Article
In 2020, Daniel P.The Abraham team used the in-situ electrochemical pool to characterize the thickness swelling of the graphite and silicon anode, quantitatively analyzed the difference in electrode swelling in different proportions of silicon carbon mixing, and analyzed several factors affecting the in-situ swelling test for the reference of subsequent researchers.
2. Sample Preparation and Testing Device
Different test electrode sheets were prepared for the assembly of the buckle and in-situ electrolytic pools following the proportions in Table 1. Different pole sheet swelling thickness determination was performed using the apparatus shown in FIG. 1.
Table 1. Composition and performance parameters of the test electrodes.
Figure 1. Schematic diagram of the electrochemical swelling test structure.
3.Interpretation of Result
The swelling test curve of the negative electrode sheet with 15% mass percentage of silicon in Figure 2 shows that the maximum swelling thickness of the cycle increases significantly, and the irreversible swelling thickness of the complete lithium stripping state of each lap is also gradually increasing, which may come from the SEI membrane and the degree of negative electrode lithium embedding.
Figure 2. Swelling curve of the silicon carbon composite anode
As can be seen from the swelling and voltage rewiring of different proportions of silicon carbon negative poles in FIG. 3, the thickness swelling for the pure graphite negative poles mainly occurs in the first and third platforms, whereas the expansion ratio of the second platform is only 1%, and there will be a certain degree of thickness shrinkage at high capacity lithium reduction. For the pure silicon negative electrode, the maximum swelling thickness reaches almost 300%, the irreversible swelling thickness is also close to 50%, and after a circle of charge and discharge, the battery capacity decay degree is also the largest.
Figure 3. potentials and swelling curves of different proportions of silicon carbon composite negative poles
From the differential capacity curve of Figure 4, it can be seen that the pure graphite electrodes have three obvious reaction peaks when embedded in lithium, while the two lithium silicon alloys of the pure silicon electrodes are higher than the graphite. When the silicon carbon is mixed, it shows a mixture of the two materials, but the reaction summit of the lithium silicon alloy is weaker.
Figure 4. Differential capacity curves of different proportions of silicon carbon composite negative poles
Figure 5 shows that when different proportions of silicon carbon are combined, the specific capacity of the electrode is gradually increased, and the maximum swelling thickness corresponding to the electrode is full, but not linearly. Usually, when the content of silicon is less than 30%, the swelling thickness of the electrode is smaller, but after more than 30%, the slope of the thickness swelling increases significantly.
Figure 5. The specific capacity and maximum swelling ratio of different proportions of silicon carbon composite anode.
Figure 6 shows the three factors affecting the electrochemical swelling test device: electrolyte, test pressure, electrode porosity, the author found that because the sealing of the electrochemical in-situ pool is not good, so the potential capacity curve of the two structures will differ, but can be weakened by selecting LiFSi electrolyte and reducing the test ratio, and the porosity of the electrode will also affect the trend of electrode swelling thickness in the process of lithium removal. In addition to the above three, the authors considered the displacement sensor drift, test pressure applied, and the effect of gas production on swelling thickness measurements.
Figure 6. Three influencing factors of the electrochemical swelling test device: electrolyte, test pressure, and electrode porosity.
4. Summarize
In this paper, the thickness swelling of graphite and silicon anode is characterized by the in-situ electrochemical pool to quantitatively analyze the difference in electrode swelling in different proportions of silicon carbon mixing, mainly with the following conclusions:
①Pure graphite electrodes swelling by ~19%, while pure silicon electrodes swelling by ~300%.
②The proportion of polar wafer swelling containing silicon will be affected by the porosity of the polar wafer design and may be higher than expected.
③For different ratios of the silicon carbon anode, the swelling is nonlinear.
④Limiting the capacity of the silicon carbon anode can adjust the maximum swelling ratio.
⑤The lithium embedded depth of the pole is better than the inhibition of swelling.
⑥The proportion of pole sheet swelling with low porosity should be larger than those with high porosity.
⑦Researchers should note the effects of several factors mentioned when using the electrochemical in-situ swelling device.
5. IEST Related Test Equipment Recommended:
SWE Series In-situ Swelling Analysis System
1.A Variety of Cell in-situ Characterization Methods (stress&swelling thickness): the swelling thickness and swelling force of the cell charging and discharge process are measured at the same time, so as to quantify the changes of cell swelling thickness and swelling force.
2.More Refined and Stable Test System: Using a highly stable and reliable automatic adjustment platform, equipped with a high precision thickness measurement sensor and a pressure regulation system, the relative thickness measurement resolution is 0 m, to realize the long cycle monitoring of the long-term charging and discharge process of cells.
3.Diversity of Environmental Control and Test Functions: SWE series equipment can adjust the temperature of the charge and discharge environment to help study the swelling behavior of cell cells under high and low temperature; In addition to conventional thickness and pressure test, cell swelling force, compression modulus and compression rate can be tested.
6. Reference
Andressa Y.R.Prado, Marco-Tulio F.Rodrigues, Stephen E.Trask, Leon Shaw and Daniel P.Abraham, Electrochemical Dilatometry of Si-Bearing Electrodes: Dimensional Changes and Experiment Design, Journal of The Electrochemical Society, 167(2020) 160551.
Subscribe Us
Contact Us
If you are interested in our products and want to know more details, please leave a message here, we will reply you as soon as we can.